A new generation of modular, high precision Druck
pressure controllers & indicators, designed for ATE
and test bench applications.
Features include
The Test Program option provides a facility for creating,
storing and executing numerous test procedures within the
instrument itself. This is particularly useful for longer, more
repetitive and laborious procedures requiring manual inputs
for rapid prototyping, manufacturing and life cycle testing.
Test Programs can also be transferred to a PC using a mass
storage device for further editing, and copied back from the
mass storage device to the instrument.
Leak Test applies a test pressure(s) to an external system
connected to the instrument to determine the magnitude of
pressure variations due to leaks. This application sets the test
pressure and a dwell time to eliminate potential adiabatic
effects at the test pressure and the leak test time period.
On completion, the display shows the Start Pressure, End
Pressure, Pressure Change and Leak Rate.
Switch Test automates the testing of pressure switch devices.
Following the test, the display shows the pressure at which
the contacts open and closes, and the switch hysteresis.
Switch Test Task can also be set to loop (n) times to exercise
a switch or capture switch toggle max, min and average
values.
The analog output option can be programmed via the setup menu screen to output a signal proportional to the instrument range
selected. This allows the instrument to interface with PC or
PLC I/O cards, remote displays, chart recorders or other data
logging equipment.
Burst Test is a brand new application for the PACE
Series designed primarily for manufacturers of
pressure rupture discs, other safety critical devices,
even medical products where it is important to
accurately measure the exact point at which the
elastic limit or point at which the device will rupture or
burst will occur.